Contamination Check

If contamination of any compound is detected, this method can quickly check to see if it is coming from your SPE-03 system, pressurized air/nitrogen supply, or solvent.

  1. Run a "clean sys" method using your final collection solvent as solvent 1.

  2. Place 8 new SPE cartridges on the SPE-03

  3. Place 8 fraction tubes in the fraction rack.

  4. Run the following method

Action

Inlet 1

Flow

Volume

Description

Elute W2

MeOH

10

5

Clean cartridges with 5mL solvent

Air-Purge W2

Air

10

5

Purge solvent from cartridges

Collect 1

MeOH

5

8

Collect 8mL of solvent through solvent and cartridge lines

Air-Purge 1

Air

10

5

Purge remaining solvents into fractions. This is your 1st set of fractions to analyze for contamination from solvent and cartridge lines

Wait

Manual Resume

System will pause, replace fraction tubes with new ones

Clean

MeOH

10

8

Deliver 8mL of solvent to sample lines

Collect

Sample

5

12

Collect solvent from sample lines to analyze for contamination from sample lines, extra 4mL as air purge. This is your 2nd set of fractions to analyze for additional contamination from sample lines.

Wait

Manual Resume

System will pause, replace fraction tubes with new ones.

Blow N2

Time-based

5 min

Dry cartridges with N2 for 5 mins.

Collect 1

MeOH

5

8

Collect 8mL of solvent through solvent and cartridge lines to analyze for contamination from nitrogen source.

Air-purge 1

Air

10

5

Purge remaining solvents into fractions. This is your 3rd set of fractions to analyze for contamination from nitrogen supply.

  1. Directly pipet 8 mL of MeOH from the solvent bottle on the SPE-03 into a fraction tube as control.

  2. Evaporate and analyze all 3 sets of fractions and 1 control.

  3. Compare the contamination levels from the 3 locations as well as control to see which was the main contributor.