LUMERA Site Preparation
LUMERA 8-Channel Cleanup Station
Site Preparation Guide
General Information
This guide helps you prepare you laboratory facility for the installation of the LUMERA 8-Channel cleanup station. Proper site preparation is critical to the successful operation of the system.
This site preparation guide applies for the following system part numbers:
LUM-92T-UVP
LUM-96T-UVP
LUM-92T
LUM-96T
Customer Support
If you have questions about this guide or preparing your site, contact support@promochrom.com
Safety Advisories
Work Area Requirements
The following figure shows a typical Lumera automated SPE system.

The system should be placed on a stable laboratory bench or fume hood with an open space of at least 80 cm wide x 50 cm deep (31.5" x 19.7") to provide access to the front and side of the instrument for fraction collection and sample loading.
There should be space underneath the bench for the placement of waste containers. The instrument must be elevated well above the waste container to allow for proper drainage.
Proper ventilation is required if toxic and volatile buffers are to be used.
Power Requirements
Refer to the following power requirements when preparing your laboratory
Electrical Safety
Follow all local electrical safety requirements in preparing you laboratory
Power Source
All systems require a dedicated power source. The receptacles from this power source should be accessible to the operator for easy disconnection from the source. If you are unsure what power source comes with your system, please contact support@promochrom.com.
Plug/Receptacle Types
Laboratory Location | Plug Type |
|---|---|
Canada, United States, Mexico, Japan | NEMA 5-15P ![]() |
Other | CEE 7/VII "Schuko" 1.6A ![]() |
Environmental Requirements
Air Quality
Ensure that the laboratory is not exposed to excessive dust
Humidity
Ensure that the relative humidity of the laboratory is between 20% and 80%, noncondensing
Air flow
Ensure that air flow from heating or air-conditioning diffusers is not directed on the system.
Temperature
The ambient temperature must be from 2 to 40 C (35.6 to 104 F).
Waste Collection Requirements
The waste management system is a gravity-driven drainage system that effectively collects and removes any waste.
There should be space underneath the bench for the placement of waste container. The instrument must be elevated well above the waste container to allow for proper drainage.
To maintain proper drainage and leak control, ensure that the system is on a stable, level, surface. The waste tubing must continuously be on a downwards angle to avoid flooding the system.

Items Customers Must Supply
Laboratory Glassware and Accessories
Buffer bottles with GL45 caps
15 mL or 50 mL conical bottom centrifuge tubes for fraction collection
Protein purification cartridges for method validation
500 mL squeeze bottle for cleaning
Clean sample bottles
4L or larger waste container
Solvents & Reagents
DI Water, HPLC grade or better
Other reagents and solvents required for method validation