Preventative Maintenance

Beginning/end of day

Between extractions

Every ~200 extractions

Every ~400 extractions

Yearly

Procedure

Line-purging and re-priming

X

See "Daily Line Purge Procedure" article

Clean System

X

X

Flush all active lines with cleaning buffers (IPA/EtOH/MeOH). Run the "clean sys" method

Deep Clean with 0.5-1.0M NaOH

X

X

Run a "clean sys" using alkaline wash buffer

Pump Maintenance

X

X

See "Syringe Pump Cleaning" Guide

Valve Cleaning

X

X

See "Valve Cleaning" Guide

UV Flow-cell Cleaning

X

If you observe a high baseline or abnormal absorbance signal on the UV touchscreen, the flow cell cuvettes can be removed from the modules for cleaning.

Preventative Maintenance Kit installation

X

Reach out to support@promochrom.com for more information