Put the syringe in maintenance position

Move the syringe to a service position to make replacement easier.

  1. On the robot, create a rinse program.
  2. In the program parameters, change the syringe positioning so it moves to the maintenance position. You must switch to Advanced user to see this parameter.
  3. Run the program.

When the program finishes, the piston does not go all the way up, which makes the syringe easier to replace.